Clinical Performance of the CTGCTV2 Assay for the BD COR System for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae From Extragenital Specimens.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Van Der Pol Barbara, Ippolito Jordan, Gao Rachel, Passow Courtney, Christensen Katherine

Аннотация

BACKGROUND: Extragenital (oropharyngeal and anorectal) Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (or gonococci [GC]) infections are underdiagnosed. Here we report sensitivity and specificity estimates that were obtained from the multiplex, polymerase chain reaction-based BD CTGCTV2 (CTGCTV2) assay, which simultaneously detects CT and GC from extragenital specimens on the high-throughput BD COR (COR) system.

METHODS: Testing on the index assay used remnant (collected during a predicate, registrational trial) oropharyngeal (n = 2318) or anorectal (n = 2292) specimens, stored in PreservCyt ThinPrep liquid-based cytology media, from sexually active individuals (≥15 years of age). A composite comparator algorithm was used to define infectious status in the predicate study, which consisted of a 2 out of 3 agreement from Food and Drug Administration-cleared CT/GC molecular assays for a reference-positive or reference-negative result. Results obtained from the CT/GC assay on the COR were compared with the predefined composite comparator algorithm.

RESULTS: Testing for CT on COR had an estimated sensitivity of 100% (95% confidence interval, 86.2%-100%) and 97.7% (93.5%-99.2%) from oropharyngeal and anorectal specimens, respectively. GC testing on COR resulted in sensitivity estimates of 92.8% (85.8%-96.5%) and 95.8% (89.7%-98.4%) from oropharyngeal and anorectal specimens, respectively. Specificity estimates for CT were 99.8% (99.5%-99.9%) and 99.4% (99.0%-99.7%), respectively. For GC, specificity estimates were 99.5% (99.1%-99.7%) and 99.8% (99.5%-99.9%) for oropharyngeal and anorectal specimens, respectively.

CONCLUSIONS: CTGCTV2 performance on COR was accurate when testing oropharyngeal and anorectal specimens.