Validation of a new wearable device for type 3 sleep test without flowmeter.

СтатьяВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Contini Mauro, Sarmento Antonio, Gugliandolo Paola, Leonardi Alessandra, Longinotti-Buitoni Gianluigi, Minella Camilla, Vignati Carlo, Mapelli Massimo, Aliverti Andrea, Agostoni Piergiuseppe

Аннотация

BACKGROUND: Ventilation monitoring during sleep is performed by sleep test instrumentation that is uncomfortable for the patients due to the presence of the flowmeter. The objective of this study was to evaluate if an innovative type 3 wearable system, the X10X and X10Y, is able to correctly detect events of apnea and hypopnea and to classify the severity of sleep apnea without the use of a flowmeter.

METHODS: 40 patients with sleep disordered breathing were analyzed by continuous and simultaneous recording of X10X and X10Y and another certified type 3 system, SOMNOtouch, used for comparison. Evaluation was performed in terms of quality of respiratory signals (scores from 1, lowest, to 5, highest), duration and classification of apneas, as well as identification and duration of hypopneas.

RESULTS: 580 periods were evaluated. Mean quality assigned score was 3.37±1.42 and 3.25±1.35 for X10X and X10Y and SOMNOtouch, respectively. The agreement between the two systems was evaluated with grades 4 and 5 in 383 out of 580 cases. A high correlation (r2 = 0.921; p<0.001) was found between the AHI indexes obtained from the two systems. X10X and X10Y devices were able to correctly classify 72.3% of the obstructive apneas, 81% of the central apneas, 61.3% of the hypopneas, and 64.6% of the mixed apneas when compared to SOMNOtouch device.

CONCLUSION: The X10X and X10Y devices are able to provide a correct grading of sleep respiratory disorders without the need of a nasal cannula for respiratory flow measurement and can be considered as a type 3 sleep test device for screening tests.

Предметные рубрики MeSH