Alignment and articular orientation of lower limbs: manual vs computer-aided measurements on digital radiograms.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Rozzanigo Umberto, Pizzoli Andrea, Minari Chiara, Caudana Roberto
Аннотация
PURPOSE: To compare the manual measurements of lower limbs on digital images with those obtained with dedicated software.
MATERIALS AND METHODS: Forty patients with a clinical suspicion of lower limb deformity were enrolled. Eighty digital radiographs were produced with a remote-controlled radiography system (Philips Omnidiagnost). The measurements were taken separately by an Orthopaedic Surgeon and by a Radiologist, by hand and with the aid of software, respectively. Five parameters were assessed: femoral length, tibial length, distal-medial femoral angle, proximal-medial tibial angle and tibial-femoral angle. The statistical analysis of the comparison was based on Student's t-test. The inter-observer variability of the methods, manual and computer-aided, was evaluated with Fisher's F-test on a sample of measurements (20 lower limbs), taken by 5 different Orthopaedic Surgeons and Radiologists, respectively.
RESULTS: There were no statistically significant differences between the measurements taken with the manual and computer-aided methods (p<0.05). The overall reproducibility of both methods was similar; conversely, the separate evaluation of angles and lengths showed that the computer-aided method was less variable in the measurement of angles and a little more variable in the measurement of lengths than the manual method.
CONCLUSIONS: The computer-aided evaluation of the alignment and articular orientation parameters of lower limbs is as accurate and reliable as the traditional manual method, but is faster and allows better-quality images.
MATERIALS AND METHODS: Forty patients with a clinical suspicion of lower limb deformity were enrolled. Eighty digital radiographs were produced with a remote-controlled radiography system (Philips Omnidiagnost). The measurements were taken separately by an Orthopaedic Surgeon and by a Radiologist, by hand and with the aid of software, respectively. Five parameters were assessed: femoral length, tibial length, distal-medial femoral angle, proximal-medial tibial angle and tibial-femoral angle. The statistical analysis of the comparison was based on Student's t-test. The inter-observer variability of the methods, manual and computer-aided, was evaluated with Fisher's F-test on a sample of measurements (20 lower limbs), taken by 5 different Orthopaedic Surgeons and Radiologists, respectively.
RESULTS: There were no statistically significant differences between the measurements taken with the manual and computer-aided methods (p<0.05). The overall reproducibility of both methods was similar; conversely, the separate evaluation of angles and lengths showed that the computer-aided method was less variable in the measurement of angles and a little more variable in the measurement of lengths than the manual method.
CONCLUSIONS: The computer-aided evaluation of the alignment and articular orientation parameters of lower limbs is as accurate and reliable as the traditional manual method, but is faster and allows better-quality images.
Предметные рубрики MeSH
AdolescentAdultAgedAged, 80 and overBody Weights and MeasuresChildFemaleFemurHumansImage Processing, Computer-AssistedLower ExtremityLower Extremity Deformities, CongenitalMaleMiddle AgedObserver VariationRadiographic Image EnhancementRadiographic Image Interpretation, Computer-AssistedReproducibility of ResultsSoftwareTibia