Accuracy of pulse oximeters at low oxygen saturations in children with congenital cyanotic heart disease: An observational study.

Kim Eun-Hee, Lee Ji-Hyun, Song In-Kyung, Kim Hee-Soo, Jang Young-Eun, Yoo Seokha, Kim Jin-Tae

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Pulse oximetry overestimates arterial oxygen saturation (SaO2 ) at less than 90% saturation in cyanotic children. The Masimo Blue sensor (Masimo Corp., Irvine, CA) is a pulse oximetry sensor developed for use in children with cyanosis. However, there remains a lack of research in actual clinical practice.

AIMS: We evaluated the intraoperative performance of three different pulse oximeters to measure oxyhemoglobin saturation (SpO2 ) at low saturations in pediatric patients with cyanotic heart disease and the influence of clinical variables (SaO2 , hemoglobin concentration, perfusion index, and weight) on the accuracy of the sensors.

METHODS

This prospective observational study compared SpO2 measured using three pulse oximeters (Masimo Blue [Masimo Corp., Irvine, CA]; Masimo LNCS, and Nellcor [Medtronic, Dublin, Ireland]) at selected SaO2 ranges (≥85%, 75%-84%, 60%-74%, and < 60%). Accuracy was evaluated according to bias and Bland-Altman analysis with appropriate correction for multiple measurements. Relationships between bias and clinical variables were assessed using a generalized estimating equation.

RESULTS

Two hundred and fifty-eight samples were analyzed. The mean overall bias (limits of agreement) of Masimo Blue, Masimo LNCS, and Nellcor sensor was -5.3 (-20.9 to 10.3%), -7.4 (-21.9 to 7.1%), and -7.4 (-22.5 to 15.1%), respectively. However, there was no difference in bias among the three sensors at SaO2 <60%. Generalized estimating equation showed that SaO2 value was associated with bias of all sensors. Perfusion index affected the bias of Blue sensor and LNCS sensor, and patients' weight was associated with bias of Nellcor sensor.

CONCLUSION

Masimo blue sensor demonstrated overall lower bias compared to the other two sensors. However, the accuracy of all sensors was similarly poor at SaO2 less than 60%. Bias was influenced by SaO2 , perfusion index, and body weight.

Для цитирования: Kim Eun-Hee, Lee Ji-Hyun, Song In-Kyung, Kim Hee-Soo, Jang Young-Eun, Yoo Seokha, Kim Jin-Tae. Accuracy of pulse oximeters at low oxygen saturations in children with congenital cyanotic heart disease: An observational study.. Paediatric anaesthesia. 2019. DOI: 10.1111/pan.13642.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.