Validation of Compressed Sensing Accelerated Magnetization Prepared Rapid Acquisition Gradient Echo for Fast and Reliable Brain Volume Measurement.

Lee Jung Bin, Ahn Hye Shin, Joo Leehi, Yim Younghee

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

Objectives

To assess the clinical viability of compressed sensing magnetization, we prepared a rapid acquisition gradient echo (CS-MPRAGE) for brain volume measurement by comparing its scan time and image quality with standard MPRAGE.

Methods

In this retrospective study, we analyzed a total of 40 morphologically normal MRIs from relatively young subjects (mean age 27.03 ± 5.02, range 19~40, F:M = 33:7) who underwent both CS-MPRAGE and standard MPRAGE. MRI was performed with a 3T MR scanner using a 64-channel head coil. Volumetric data analysis was conducted using commercial AI-powered quantification software. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNRgray matter/white matter) were measured and compared. Qualitative analysis based on overall image quality, deep gray matter delineation, artifacts, and gray-white matter differentiation was rated on a five-point visual scoring system. Statistical analyses included paired t-tests, intraclass correlation coefficients (ICCs) and Pearson correlation.

Results

CS-MPRAGE had a significantly shorter scan time than standard MPRAGE (1:46 vs. 5:20 min; p < 0.001). Whole-brain volume was similar (standard-MPRAGE, 1176.89 ± 105.36; CS-MPRAGE, 1181.14 ± 105.62 mL; r = 0.988; ICC, 0.9754-0.9931). Gray matter and regional lobe volumes were comparable, while the white matter volume was slightly larger with CS-MPRAGE. CNR of the gray-white matter was significantly higher with CS-MPRAGE (left, 30.45 ± 11.21 vs. 49.53 ± 32.71; right, 30.73 ± 11.97 vs. 46.89 ± 30.40; p = 0.0001) while SNR were similar across key brain regions.

Conclusions

CS-MPRAGE provides high-quality 3D images and reliable volume data with significantly reduced acquisition time and comparable image quality.

Для цитирования: Lee Jung Bin, Ahn Hye Shin, Joo Leehi, Yim Younghee. Validation of Compressed Sensing Accelerated Magnetization Prepared Rapid Acquisition Gradient Echo for Fast and Reliable Brain Volume Measurement.. Tomography (Ann Arbor, Mich.). 2026. DOI: 10.3390/tomography12090121.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.