DOI 10.23736/s2724-5276.21.06021-7

Медицинская библиотека · Minerva pediatrics · 2021

How can stretched ductus diameter be predicted from angiographic diameter?

Kim Hanna, Song Jinyoung, Choi Gwang-Jun, Huh June, Kang I-Seok

СтатьяПеревод в очереди

Аннотация

BACKGROUND

An appropriate size of device for patent ductus arteriosus (PDA) could be chosen by a stretched PDA size. We propose prediction of stretched size from intact PDA size.

METHODS

A total of 361 patients was enrolled. Intact size was measured on angiography before closure, and stretched size was the narrowest diameter immediately after device implantation.

RESULTS

The median patient age was 2.8 years. Intact diameter and stretched diameter were 3.6±1.4 mm and 5.1±1.3 mm. The difference and difference ratio were 1.5±0.8 mm and 51.3±38.1% and had negative linear correlations with age (P<0.001). Patients were divided into four groups: <2 years old (159), 2-15 years old (68), 15-40 years old (72), and ≥40 years old (62). Among groups, difference and difference ratio were significantly different. Stretched diameter was inferred by the following formulas: stretched diameter = 0.660 × intact diameter + 2.791 (<2 years old); = 0.971×intact diameter + 0.058 × age + 1.131 (2-15 years old); = 0.790 × intact diameter + 2.263 (15-40 years old); and = 0.837 × intact diameter + 1.353 × BSA - 0.096 (≥40 years old).

CONCLUSIONS

Stretched PDA diameter can be predicted from intact diameter and might be helpful for safe PDA closure especially in small infants.

Для цитирования: Kim Hanna, Song Jinyoung, Choi Gwang-Jun, Huh June, Kang I-Seok. How can stretched ductus diameter be predicted from angiographic diameter?. Minerva pediatrics. 2021. DOI: 10.23736/s2724-5276.21.06021-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.