Validation of Omron HBP-1300 professional blood pressure monitor based on auscultation in children and adults.
СтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Meng Linghui, Zhao Di, Pan Yan, Ding Wenqing, Wei Qing, Li Hua, Gao Pingjin, Mi Jie
Аннотация
UNLABELLED: BACKGROUnD: To determine whether the professional Omron HBP-1300 blood pressure (BP) monitor meets American Association for the Advancement of Medical Instrumentation (AAMI) accuracy standards in Chinese children and adults.
METHOD: According to the AAMI protocol, simultaneous auscultatory measurements by two observers using a mercury manometer were obtained in participants using the Omron HBP-1300. Triple measurements were obtained after a minimum 5-min rest with a 1-min interval between adjacent measurements.
RESULTS: A total of 85 participants submitted to 255 doctor-measured BP and 255 successful professional monitor readings. The initial auscultation systolic BP was <100 mmHg in 25 participants (29.4%), 100-160 mmHg in 53 participants (62.4%), and >160 mmHg in seven participants (8.2%). All of the simultaneous measurements agreed to within ± 10 mmHg, while 95% agreed to within ± 4 mmHg for both systolic and diastolic BP, and the consistency between two observers was satisfactory. The difference between the devices was -1.3 ± 3.6 mmHg for systolic BP and 0.7 ± 3.8 mmHg for diastolic BP and by AAMI method 1, which met this guideline. The average difference between two devices by AAMI method 2 was 1.4 ± 3.2 mmHg for systolic BP and 1.0 ± 3.9 mmHg for diastolic BP, which met this guideline.
CONCLUSION: The professional BP monitor Omron HBP-1300 is desirable for measuring the BP for Chinese children and adults.
METHOD: According to the AAMI protocol, simultaneous auscultatory measurements by two observers using a mercury manometer were obtained in participants using the Omron HBP-1300. Triple measurements were obtained after a minimum 5-min rest with a 1-min interval between adjacent measurements.
RESULTS: A total of 85 participants submitted to 255 doctor-measured BP and 255 successful professional monitor readings. The initial auscultation systolic BP was <100 mmHg in 25 participants (29.4%), 100-160 mmHg in 53 participants (62.4%), and >160 mmHg in seven participants (8.2%). All of the simultaneous measurements agreed to within ± 10 mmHg, while 95% agreed to within ± 4 mmHg for both systolic and diastolic BP, and the consistency between two observers was satisfactory. The difference between the devices was -1.3 ± 3.6 mmHg for systolic BP and 0.7 ± 3.8 mmHg for diastolic BP and by AAMI method 1, which met this guideline. The average difference between two devices by AAMI method 2 was 1.4 ± 3.2 mmHg for systolic BP and 1.0 ± 3.9 mmHg for diastolic BP, which met this guideline.
CONCLUSION: The professional BP monitor Omron HBP-1300 is desirable for measuring the BP for Chinese children and adults.