MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/s0090-4295(03)00010-4

Медицинская библиотека · Urology · 2003

Causes for variability in repeated pressure-flow measurements.

Kranse Ries, van Mastrigt Ron

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To study the causes for the variation between measurements, which is considerable, in maximal flow rate (Qmax) and the associated detrusor pressure (PdetQmax). Because of the central role of Qmax and PdetQmax in the diagnosis of bladder outlet obstruction, this is reason for concern.

METHODS

Spectral analysis was carried out on two consecutive urodynamic measurements in 131 patients. The parameters for bladder outlet resistance and bladder contraction strength were determined, and difference plots were made to study the systematic variations. Logistic regression analysis was used to study whether the differences represent true changes of the function of the lower urinary tract.

RESULTS

Signal components in the detrusor pressure and the flow rate signal with frequencies of 1 Hz or greater may be considered noise. Filtering out these frequencies changes the estimates of Qmax and PdetQmax, but not the between-measurement difference in them. Bladder contractility and bladder outlet resistance were systematically lower in the second measurement. Both the systematic and nonsystematic between-measurement variations were statistically significant predictors for postvoid residual urine volume.

CONCLUSIONS

The nonsystematic between-measurement variability in Qmax and PdetQmax apparently reflects true variability in the physiologic state of the bladder outlet. It therefore does not discredit the pressure-flow study as the preferred method in the diagnosis of bladder outlet obstruction. Rather, the pressure-flow study is the only currently available method to study and quantify the apparent within-patient variability in bladder outlet resistance and bladder contractility.

Для цитирования: Kranse Ries, van Mastrigt Ron. Causes for variability in repeated pressure-flow measurements.. Urology. 2003. DOI: 10.1016/s0090-4295(03)00010-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.