DOI 10.1109/tbme.2022.3227015

Медицинская библиотека · IEEE transactions on bio-medical engineering · 2023

Approach to Quantify Eye Movements to Augment Stroke Diagnosis With a Non-Calibrated Eye-Tracker.

Abul Hassan Mohamed, Aldridge Chad M, Zhuang Yan, Yin Xuwang, McMurry Timothy, Rohde Gustavo K, Southerland Andrew M

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

UNLABELLED: Automated eye-tracking technology could enhance diagnosis for many neurological diseases, including stroke. Current literature focuses on gaze estimation through a form of calibration. However, patients with neuro-ocular abnormalities may have difficulty completing a calibration procedure due to inattention or other neurological deficits.

OBJECTIVE

We investigated 1) the need for calibration to measure eye movement symmetry in healthy controls and 2) the potential of eye movement symmetry to distinguish between healthy controls and patients.

METHODS

We analyzed fixations, smooth pursuits, saccades, and conjugacy measured by a Spearman correlation coefficient and utilized a linear mixed-effects model to estimate the effect of calibration.

RESULTS

Healthy participants (n = 18) did not differ in correlations between calibrated and non-calibrated conditions for all tests. The calibration condition did not improve the linear mixed effects model (log-likelihood ratio test p = 0.426) in predicting correlation coefficients. Interestingly, the patient group (n = 17) differed in correlations for the DOT (0.844 [95% CI 0.602, 0.920] vs. 0.98 [95% CI 0.976, 0.985]), H (0.903 [95% CI 0.746, 0.958] vs. 0.979 [95% CI 0.971, 0.986]), and OKN (0.898 [95% CI 0.785, 0.958] vs. 0.993 [95% CI 0.987, 0.996]) tests compared to healthy controls along the x-axis. These differences were not observed along the y-axis.

SIGNIFICANCE: This study suggests that automated eye tracking can be deployed without calibration to measure eye movement symmetry. It may be a good discriminator between normal and abnormal eye movement symmetry. Validation of these findings in larger populations is required.

Для цитирования: Abul Hassan Mohamed, Aldridge Chad M, Zhuang Yan, Yin Xuwang, McMurry Timothy, Rohde Gustavo K, Southerland Andrew M. Approach to Quantify Eye Movements to Augment Stroke Diagnosis With a Non-Calibrated Eye-Tracker.. IEEE transactions on bio-medical engineering. 2023. DOI: 10.1109/tbme.2022.3227015.