Strategies for Computed Tomography Radiation Dose Reduction in Pediatric Neuroimaging.

Albert Gregory W, Glasier Charles M

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Radiation exposure from diagnostic imaging is a significant concern, particularly in the care of pediatric patients. Computed tomography (CT) scanning is a significant source of radiation.

OBJECTIVE

To demonstrate that diagnostic quality CT images can be obtained while minimizing the effective radiation dose to the patient.

METHODS

In this retrospective cross-sectional study, noncontrast head CT scan data were reviewed, and indications for scans and estimated radiation dose delivered were recorded. The estimated effective radiation dose (EERD) for each CT protocol was reviewed.

RESULTS

We identified 251 head CT scans in a single month. Of these, 96 scans were using a low-dose shunt protocol with a mean EERD of 0.82 mSv. The remaining 155 scans were performed using the standard protocol, and the mean EERD was 1.65 mSv. Overall, the EERD was minimized while maintaining diagnostic scan quality.

CONCLUSION

Although replacing a CT with magnetic resonance imaging is ideal to completely avoid ionizing radiation, this is not always practical or preferred. Therefore, it is important to have CT protocols in place that minimize radiation dose without sacrificing diagnostic quality. The protocols in place at our institution could be replicated at other academic and community hospitals and imaging centers.

Для цитирования: Albert Gregory W, Glasier Charles M. Strategies for Computed Tomography Radiation Dose Reduction in Pediatric Neuroimaging.. Neurosurgery. 2015. DOI: 10.1227/neu.0000000000000764.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.