MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/labmed/lmac124

Медицинская библиотека · Laboratory medicine · 2023

Establishment of Review Criteria Coordinating With the Automated Digital Cell Morphology Identification System in a Specialized Women's and Children's Hospital.

Wang Yuefang, Ye Lei, Chen Lan, Chen Qi, Zhang Xia, Dai Qingkai, Peng Luyun, Lai Chunqi, Zhang Ge

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

We aimed to establish appropriate review criteria for blood cell analysis in a specialized women's and children's hospital. Also, the CellaVision DI-60, was developed as one of the automated digital cell morphology analyzer, we evaluated if it was shown to be most effective under the certain review criteria.

METHODS

A total of 2890 blood samples were detected to optimize the previously established review criteria for women and children with the Sysmex XE-2100. A total of 623 samples were used to validate the criteria.

RESULTS

The microscopic-review rate based on the initial review criteria was 51.0%. After optimization, it was reduced to 17.3% and the false-negative rate was 3.85%. There was > 80% consistency between manual review results and CellaVision DI-60 preclassification when samples triggered the platelet- or red cell-related rules. The sensitivity for abnormalities (immature granulocytes, nucleated red blood cells) of reclassification was 90% to 100% and the false-negative rate was < 5%. However, direct microscopic review was required when the "Blasts/AbnLympho?" and "Atypical Lympho?" flags were triggered.

CONCLUSION

Specialized review criteria are needed for women and children. An automated morphology identification system might help to improve the review criteria.

Для цитирования: Wang Yuefang, Ye Lei, Chen Lan, Chen Qi, Zhang Xia, Dai Qingkai, Peng Luyun, Lai Chunqi, Zhang Ge. Establishment of Review Criteria Coordinating With the Automated Digital Cell Morphology Identification System in a Specialized Women's and Children's Hospital.. Laboratory medicine. 2023. DOI: 10.1093/labmed/lmac124.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.