Design and Validation of a 3D Printed Cranio-Facial Simulator: A Novel Tool for Surgical Education.
Аннотация
OBJECTIVE
To assess the ability of current 3D printing technology to generate a craniofacial bony and soft tissue anatomical model for use in simulating the performance of a fronto-orbital advancement (FOA) osteotomy and then to further assess the value of the model as an educational tool.
DESIGN
Anatomic models were designed with a process of serial anatomic segmentation/design, 3D printing, dissection, and device refinement. A validation study was conducted with 5 junior and 5 senior plastic surgery residents. The validation study incorporated a multiple-choice Knowledge Assessment test (KA), an Objective Structured Assessment of Technical skills (OSATs), a Global Rating Scale (GRS) and a Michigan Standard Simulation Experience Scale (MiSSES). We compared the scores of both the junior and senior residents and compared junior resident scores, before and after viewing a lecture/demonstration.
RESULTS
MiSSES showed high face validity with a score of 85.1/90, signifying high satisfaction with the simulator learning experience. Simulation and the lecture/demonstration improved the junior resident average KA score from 5.6/10 to 9.6/10 (P = .02), OSATs score from 32.4/66 to 64.4/66 (P < .001) and GRS score from 13.9/35 to 27.5/35 (P < .001). The senior residents OSATs score of 56.3/66 was higher than the pre-lecture juniors (32.4/66) (P < .001), but lower than the post-lecture juniors (64.4/66) (P < .001).
CONCLUSION
We have successfully fabricated a 3D printed craniofacial simulator capable of being used as an educational tool alongside traditional surgical training. Next steps would be improving soft tissue realism, inclusion of patient and disease specific anatomy and creation of models for other surgical specialties.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.