Dosimetric assessment of static and helical TomoTherapy in the clinical implementation of breast cancer treatments.

Reynders Truus, Tournel Koen, De Coninck Peter, Heymann Steve, Vinh-Hung Vincent, Van Parijs Hilde, Duchateau Michaël, Linthout Nadine, Gevaert Thierry, Verellen Dirk, Storme Guy

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND AND PURPOSE: Investigation of the use of TomoTherapy and TomoDirect versus conventional radiotherapy for the treatment of post-operative breast carcinoma. This study concentrates on the evaluation of the planning protocol for the TomoTherapy and TomoDirect TPS, dose verification and the implementation of in vivo dosimetry.

MATERIALS AND

METHODS

Eight patients with different breast cancer indications (left/right tumor, axillary nodes involvement (N+)/no nodes (N0), tumorectomy/mastectomy) were enrolled. TomoTherapy, TomoDirect and conventional plans were generated for prone and supine positions leading to six or seven plans per patient. Dose prescription was 42Gy in 15 fractions over 3weeks. Dose verification of a TomoTherapy plan is performed using TLDs and EDR2 film inside a home-made wax breast phantom fixed on a rando-alderson phantom. In vivo dosimetry was performed with TLDs.

RESULTS

It is possible to create clinically acceptable plans with TomoTherapy and TomoDirect. TLD calibration protocol with a water equivalent phantom is accurate. TLD verification with the phantom shows measured over calculated ratios within 2.2% (PTV). An overresponse of the TLDs was observed in the low dose regions (<0.1Gy). The film measurements show good agreement for high and low dose regions inside the phantom. A sharp gradient can be created to the thoracic wall. In vivo dosimetry with TLDs was clinically feasible.

CONCLUSIONS

The TomoTherapy and TomoDirect modalities can deliver dose distributions which the radiotherapist judges to be equal to or better than conventional treatment of breast carcinoma according to the organ to be protected.

Для цитирования: Reynders Truus, Tournel Koen, De Coninck Peter, Heymann Steve, Vinh-Hung Vincent, Van Parijs Hilde, Duchateau Michaël, Linthout Nadine, Gevaert Thierry, Verellen Dirk, Storme Guy. Dosimetric assessment of static and helical TomoTherapy in the clinical implementation of breast cancer treatments.. Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology. 2009. DOI: 10.1016/j.radonc.2009.07.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.