DOI 10.1016/j.medengphy.2013.11.001

Медицинская библиотека · Medical engineering & physics · 2013

Neonatal phototherapy radiometers: current performance characteristics and future requirements.

Clarkson Douglas McG, Nicol Ruth, Chapman Phillip

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

Hand held radiometers provide a convenient means of monitoring the output of neonatal phototherapy treatment devices as part of planned programs of device maintenance and output monitoring. It was considered appropriate to determine the wavelength and angular response of a selection of such meters and compare their indicated values with that derived from spectral analysis of phototherapy light sources. This was undertaken using a Bentham DMc150 double grating spectroradiometer and a series of 10nm band pass optical filters in the range 400-640 nm used in conjunction with a fiber optic light source. Specific meters investigated included a GE Biliblanket Light Meter II, a NeoBLUE radiometer and a Bio-TEK radiometer 74345 device. Comparisons were made of measurements made using the hand held meters and the Bentham DMc 150 system for a range of neonatal phototherapy treatment devices. The use of such meters is discussed in relation to applicable equipment standards and recommendations of intensive phototherapy from clinical groups such as the American Academy of Pediatrics and a specification for a spectroradiometer based measurement system is proposed.

Для цитирования: Clarkson Douglas McG, Nicol Ruth, Chapman Phillip. Neonatal phototherapy radiometers: current performance characteristics and future requirements.. Medical engineering & physics. 2013. DOI: 10.1016/j.medengphy.2013.11.001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.