Does the Length of Dental Implants Inserted in Areas of Inferior Alveolar Nerve Lateralization Interfere on Mandibular Resistance to Fracture?
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Brackmann Mariana Schaffer, Padovan Luis Eduardo Marques, Marcantonio Junior Elcio, Klüppel Leandro Eduardo
Аннотация
The aim of this study was to compare, in vitro, the mechanical resistance to vertical displacement of the mandible after osteotomy for lateralization of the inferior alveolar nerve and installation of dental implants. One hundred eighty polyurethane mandibles were equally divided into 6 groups: G1-intact hemi-mandibles (control group), G2-hemi-mandibles after osteotomy for lateralization of the inferior alveolar nerve, G3-hemi-mandibles with installation of 3 bicortical dental implants (3.75 × 13 mm), G4-hemi-mandibles with installation of 3 dental implants that did not reach the basal cortical bone (3.75 × 11 mm), G5-hemi-mandibles after osteotomy for lateralization of the inferior alveolar nerve and installation of 3 bicortical dental implants (3.75 × 13 mm) and G6-hemi-mandibles after osteotomy for lateralization of the inferior alveolar nerve and installation of 3 dental implants that did not reach the basal cortical bone (3.75 × 11 mm). The specimens were subjected to linear loading tests. The highest mean value of maximum load was found in G1 (412.36N ± 11.99), followed by G2 (396.87N ± 23.94), G3 (319.63N ± 57.28), G4 (303.34N ± 18.25), G5 (231.75N ± 63.64) and G6 (228.13N ± 20.75). Based on this data, it can be concluded that the bicorticalization (or not) of the implants was not a statistically significant risk factor for the vertical displacement of polyurethane hemi-mandibles.