Acta neurochirurgica. Supplement · 1998

Simultaneous continuous measurement of pO2, pCO2, pH and temperature in brain tissue and sagittal sinus in a porcine model.

Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Menzel M, Rieger A, Roth S, Soukup J, Peuse C, Hennig C, Molnar P, Furka I, Radke J

Аннотация

INTRODUCTION: The clinical use of brain tissue oxygen measurement in patients with severe head injury is increasing. It is important to compare the findings in brain tissue with cerebrovenous blood oximetry, to obtain normal values and to find out limitations of the method. We evaluated a newly available multisensor probe simultaneously in the brain tissue and in the sagittal sinus in a porcine animal model.

METHODS: We placed the Paratrend 7-probe (BSL, High Wycombe, UK) in the left frontoparietal white matter and measured pO2 (PtiO2), pCO2 (ptiCO2), pH and temperature while simultaneously measuring these parameters (pcvO2, pcvCO2) in the sagittal sinus in 7 pigs under general anaesthesia during oxygen enhancement.

RESULTS: The relation between oxygen increase in brain tissue and in the sagittal sinus showed a coefficient of correlation (CCmean) rmean = 0.96. The quantitative response in brain tissue was much more sensitive than in the sinus. A close correlation between pCO2 in brain tissue and sagittal sinus and the increase of the inspired oxygen was seen: CC ptiCO2 to arterial oxygen pressure (paO2) - rmean = 0.67, CC pcvCO2 to paO2 - rmean = 0.88.

CONCLUSIONS: Measuring partial oxygen pressure in brain tissue is more responsive to physiological variations, and the absolute values are more sensitive than oxygen measurement in the cerebrovenous compartment. This is important for interpreting measured values and introducing new coefficients for patient monitoring.