Use of the GORE® CARDIOFORM Septal Occluder for percutaneous closure of secundum atrial septal defects: Results of the multicenter U.S. IDE trial.

Gillespie Matthew J, Javois Alexander J, Moore Phillip, Forbes Thomas, Paolillo Joseph A, GSO Investigator Group

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To evaluate safety and efficacy of the GORE® CARDIOFORM Septal Occluder for percutaneous transcatheter closure of ostium secundum atrial septal defects.

BACKGROUND

The GORE® CARDIOFORM septal occluder is a double-disc, low profile, soft, conformable device, with distinct advantages over the GORE® HELEX® Septal Occluder.

METHODS

Subjects were enrolled in this single arm prospective study from 21 U.S. sites, and followed for 3 years. Primary endpoint was 6 month composite clinical success, comprised of technical success (implantation and retention of device), closure success (normalization of right heart size), no 30 day serious adverse events, and no device embolization or reintervention. Secondary endpoints included technical success, procedure success (technical success and ≤ 2 mm residual shunt at procedure conclusion), closure success (clinically insignificant or no residual shunt), and safety (freedom from 30 day serious adverse events and 6-month device events).

RESULTS

Between October 2012 and May 2015, 50 pivotal and 350 continued access subjects underwent attempted transcatheter GORE® CARDIOFORM Septal Occluder implantation. Median age was 6.9 years, and mean static defect diameter 9.7 ± 3.1 mm. Device placement was achieved in 93.5% (374/400). Composite clinical success was 90.2% and clinical closure success was 98.8% at 6 months. Freedom from serious adverse events was 98.3% at 30 days, with no device embolizations or reinterventions through 6 months.

CONCLUSIONS

The GORE® CARDIOFORM Septal Occluder has high composite clinical success and safety, performing well in defects ≤17 mm by stop flow stretched diameter. Single, multifenestrated, and deficient retroaortic tissue defects were well represented and successfully treated.

Для цитирования: Gillespie Matthew J, Javois Alexander J, Moore Phillip, Forbes Thomas, Paolillo Joseph A, GSO Investigator Group. Use of the GORE® CARDIOFORM Septal Occluder for percutaneous closure of secundum atrial septal defects: Results of the multicenter U.S. IDE trial.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2020. DOI: 10.1002/ccd.28814.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.