Improved ejection fraction and flow velocity estimates with use of view sharing and uniform repetition time excitation with fast cardiac techniques.

Foo T K, Bernstein M A, Aisen A M, Hernandez R J, Collick B D, Bernstein T

СтатьяПеревод в очереди

Аннотация

PURPOSE: To improve the accuracy of ventricular volume estimates and effective temporal resolution in fast cardiac acquisitions.

MATERIALS AND

METHODS

Data sets at intermediate temporal phases were generated by means of sharing of views from two temporally adjacent data sets. A simulated model was used, and studies with patients and volunteers were conducted. View sharing was implemented in both fast gradient-echo and fast phase-contrast cine acquisitions; breath holding was used when possible.

RESULTS

Uniform repetition time (TR) radio-frequency (RF) excitation allowed a better assessment of the end-diastolic ventricular volume. In addition, view sharing provided a better estimate of end-systolic ventricular volume in cases in which rapid changes in volume occurred at or about the temporal boundary of the source images. View sharing also provided a much smoother representation of dynamic cardiac motion when viewed in a cine loop.

CONCLUSION

View sharing and uniform TR RF excitation improve the accuracy of end-systolic and end-diastolic ventricular volume measurements by improving the effective temporal resolution.

Для цитирования: Foo T K, Bernstein M A, Aisen A M, Hernandez R J, Collick B D, Bernstein T. Improved ejection fraction and flow velocity estimates with use of view sharing and uniform repetition time excitation with fast cardiac techniques.. Radiology. 1995. DOI: 10.1148/radiology.195.2.7724769.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.