Validation of A&D TM-2430 upper-arm blood pressure monitor for ambulatory blood pressure monitoring in children and adolescents, according to the British Hypertension Society protocol.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Yip Gabriel Wai-Kwok, So Hung-Kwan, Li Albert Martin, Tomlinson Brian, Wong Sik-Nin, Sung Rita Yn-Tz
Аннотация
OBJECTIVE: The A&D TM-2430 ambulatory blood pressure (BP) monitor has been validated in adults but not in a young population. We sought to validate the device monitoring in children and adolescents, according to the British Hypertension Society (BHS) protocol.
METHODS: The A&D TM-2430 is an automated oscillometric upper-arm device for ambulatory BP monitoring. Nine consecutive measurements were taken in 61 children (mean age, 9.8 years; range, 5-15 years) according to the BHS criteria. Overseen by an independent supervisor, measurements were recorded by two observers blinded from each other's readings and from the device readings.
RESULTS: The mean difference ± SD between the observers and device measurements was 0.73 ± 1.64 mmHg for systolic blood pressure (SBP) and -1.23 ± 1.65 mmHg for diastolic blood pressure (DBP), respectively, with an interobserver difference of 4 mmHg. The cumulative percentages of differences within 5, 10, and 15 mmHg were 89, 95, and 98% for SBP and 67, 88, and 98% for DBP. The device achieved a grade A rating for SBP and a B grade for DBP.
CONCLUSION: The A&D TM-2430 upper-arm BP monitor has fulfilled the required BHS standards and can be recommended for measuring ambulatory BP in children and adolescent populations.
METHODS: The A&D TM-2430 is an automated oscillometric upper-arm device for ambulatory BP monitoring. Nine consecutive measurements were taken in 61 children (mean age, 9.8 years; range, 5-15 years) according to the BHS criteria. Overseen by an independent supervisor, measurements were recorded by two observers blinded from each other's readings and from the device readings.
RESULTS: The mean difference ± SD between the observers and device measurements was 0.73 ± 1.64 mmHg for systolic blood pressure (SBP) and -1.23 ± 1.65 mmHg for diastolic blood pressure (DBP), respectively, with an interobserver difference of 4 mmHg. The cumulative percentages of differences within 5, 10, and 15 mmHg were 89, 95, and 98% for SBP and 67, 88, and 98% for DBP. The device achieved a grade A rating for SBP and a B grade for DBP.
CONCLUSION: The A&D TM-2430 upper-arm BP monitor has fulfilled the required BHS standards and can be recommended for measuring ambulatory BP in children and adolescent populations.