[Dose reduction of radiographs of the pediatric pelvis for diagnosing hip dysplasia using a digital flat-panel detector system].

Сравнительное исследованиеОценочное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ludwig K, Ahlers K, Sandmann C, Gosheger G, Kloska S, Vieth V, Meier N, Heindel W

Аннотация

PURPOSE: To evaluate a possible dose reduction in pediatric pelvic radiographs in congenital hip dysplasia using a digital flat-panel system instead of a phosphor-storage system.

MATERIALS AND METHODS: During a six-month period, all pediatric patients referred for pelvic radiography for the evaluation of congenital hip dysplasia were randomly assigned to be examined by either a phosphor-storage system or a digital flat-panel system, whereby the latter system was operated with half the radiation dose. Thirty pairs of radiographs were assessed for the visibility of 16 anatomic details and for 5 orthopedic-radiographic measurements (5-point scale with 1 = excellent; three independent observers). The projection indices of Ball and Kommenda and of Tönnis and Brunken were calculated for all radiographs. The Student's t-test was used to compare the flat-panel and the phosphor-storage radiographs for observers' assessments, patients' age and projection indices.

RESULTS: In a total of 7560 observations, the scores for the visibility of anatomic details and orthopedic-radiographic measurements were respectively 2.72 and 2.64 for the flat-panel system and 2.93 and 2.79 for the phosphor-storage system. No significant differences were found between both systems (p > 0.05) and between patient age and projection indices (p > 0.05).

CONCLUSION: Pediatric pelvic radiographs can be obtained with a digital flat-panel system using half the radiation dose instead of a phosphor-storage system without sacrificing relevant information in the diagnosis of congenital hip dysplasia.