DOI 10.1161/circinterventions.118.006453
Intentional Fracture of Bioprosthetic Valve Frames in Patients Undergoing Valve-in-Valve Transcatheter Pulmonary Valve Replacement.
Аннотация
BACKGROUND
Percutaneous transcatheter pulmonary valve replacement (TPVR) has good clinical and hemodynamic outcomes in treating dysfunctional bioprosthetic valves (BPV) in the pulmonary position. Valve-in-valve therapy can further decrease the inner diameter (ID), potentially resulting in patient-prosthesis mismatch in patients with smaller BPVs.
METHODS AND
RESULTS
To evaluate feasibility and outcomes of intentional BPV fracture to enlarge the pulmonary valve orifice with TPVR, 37 patients from 13 centers who underwent TPVR with intended BPV fracture were evaluated. A control cohort (n=70) who underwent valve-in-valve TPVR without attempted fracture was evaluated. BPV was successfully fractured in 28 patients and stretched in 5 while fracture was unsuccessful in 4. A Melody valve was implanted in 25 patients with fractured/stretched frame and a Sapien (XT 3) valve in 8. Among patients whose BPV was fractured/stretched, the final ID was a median of 2 mm larger (0-6.5 mm) than the valve's true ID. The narrowest diameter after TPVR in controls was a median of 2 mm smaller ( P<0.001) than true ID. Right ventricular outflow tract gradient decreased from median 40 to 8 mm Hg in the fracture group. Cases with fracture/stretching were matched 1:1 (weight, true ID) to controls. Post-TPVR peak gradient was lower but not significant (8.3±5.2 versus 11.8±9.2 mm Hg; P=0.070). There were no fracture-related adverse events.
CONCLUSIONS
Preliminary experience shows intentional fracture of BPV frame can be useful for achieving larger ID and better hemodynamics after valve-in-valve TPVR.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.