Flexor Hallucis Longus Tendon Transfer Fixation.

Lee Jessica, Williams Chad, Lowrey Charles, Gould Greg, Markert Ronald, Laughlin Richard

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Flexor hallucis longus (FHL) tendon transfer to the calcaneus is commonly used in the surgical treatment of chronic Achilles tendinopathy. This study assesses the integrity of FHL tendon biotenodesis screw fixation with respect to 2 variables: incorporation of a terminal whipstitch and tunnel depth.

MATERIALS AND

METHODS

A total of 60 fresh-frozen cadaver FHL tendons and 28 calcanei were harvested for analysis in 4 sets of fixation constructs; 14 whipstitched tendons were compared against their nonwhipstitched paired tendon via pull-out strength load testing, and 16 tendon pairs were randomized for fixation in either a full-depth tunnel (bicortical) or a 25-mm partial tunnel (unicortical). All comparisons were carried out in native bone and synthetic models.

RESULTS

Whipstitched tendons demonstrated significantly stronger mean clinical load (253.68 vs 177.24 N, P = .008) and maximum load to failure (294.31N vs 194.57 N, P = .001) compared with the nonwhipstitched tendons in synthetic bone. There were no statistical differences in mean clinical load (200.96 vs 228.31 N, P = .63) and maximum load to failure (192.69 vs 217.74 N, P = .73) between full and partial tunnel groups. There were no significant differences found in trials carried out in cadaveric bone.

CONCLUSION

Use of a terminal whipstitch achieves greater fixation strength in FHL tendon biotenodesis transfers. Complete and partial tunnel constructs are equivocal in their pull-out strength. Data produced in a homogeneous bone substitute model demonstrate the biomechanical superiority of the whipstitch as well as the noninferiority of the partial tunnel technique.

LEVELS OF EVIDENCE: Level IIb.

Для цитирования: Lee Jessica, Williams Chad, Lowrey Charles, Gould Greg, Markert Ronald, Laughlin Richard. Flexor Hallucis Longus Tendon Transfer Fixation.. Foot & ankle specialist. 2016. DOI: 10.1177/1938640016666917.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.