Point-of-care molecular diagnosis of Mycoplasma pneumoniae including macrolide sensitivity using quenching probe polymerase chain reaction.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ishiguro Nobuhisa, Sato Rikako, Mori Toshihiko, Tanaka Hiroshi, Narita Mitsuo, Nagano Takashi, Owaku Masato, Miyajima Kensuke, Manabe Atsushi
Аннотация
OBJECTIVES: Macrolides are generally considered to be the drugs of choice for treatment of patients with Mycoplasma pneumoniae infection. However, macrolide-resistant M. pneumoniae has been emerging since about 2000. The Smart Gene® system (MIZUHO MEDY Co., Ltd., Tosu, Japan) is a novel fully automated system for detection of pathogens using the method of quantitative polymerase chain reaction (qPCR) with QProbe (QProbe PCR). The entire procedure is completed within 50 min and the size of the instrument is small (15 x 34 x 30 cm). The purpose of this study was to evaluate the usefulness of the Smart Gene® system for detection of M. pneumoniae and detection of a point mutation at domain V of the 23S rRNA gene of M. pneumoniae.
MATERIALS: Pharyngeal swab samples were collected from 154 patients who were suspected of having respiratory tract infections associated with M. pneumoniae.
RESULTS: Compared with the results of qPCR, the sensitivity and specificity of the Smart Gene® system were 98.7% (78/79) and 100.0% (75/75), respectively. A point mutation at domain V of the 23S rRNA gene was detected from 7 (9.0%) of 78 M. pneumoniae-positive samples by the Smart Gene® system and these results were confirmed by direct sequencing. The minimum inhibitory concentrations of clarithromycin among the 5 isolates of M. pneumoniae with a point mutation at domain V of the 23S rRNA gene were >64 μg/ml and those among the 33 isolates without a mutation in the 23S rRNA gene were <0.0625 μg/ml.
CONCLUSION: The Smart Gene® system is a rapid and accurate assay for detection of the existence of M. pneumoniae and a point mutation at domain V of the 23S rRNA gene of M. pneumoniae at the same time. The Smart Gene® system is suitable for point-of-care testing in both hospital and outpatient settings.
MATERIALS: Pharyngeal swab samples were collected from 154 patients who were suspected of having respiratory tract infections associated with M. pneumoniae.
RESULTS: Compared with the results of qPCR, the sensitivity and specificity of the Smart Gene® system were 98.7% (78/79) and 100.0% (75/75), respectively. A point mutation at domain V of the 23S rRNA gene was detected from 7 (9.0%) of 78 M. pneumoniae-positive samples by the Smart Gene® system and these results were confirmed by direct sequencing. The minimum inhibitory concentrations of clarithromycin among the 5 isolates of M. pneumoniae with a point mutation at domain V of the 23S rRNA gene were >64 μg/ml and those among the 33 isolates without a mutation in the 23S rRNA gene were <0.0625 μg/ml.
CONCLUSION: The Smart Gene® system is a rapid and accurate assay for detection of the existence of M. pneumoniae and a point mutation at domain V of the 23S rRNA gene of M. pneumoniae at the same time. The Smart Gene® system is suitable for point-of-care testing in both hospital and outpatient settings.
Предметные рубрики MeSH
AdolescentAdultAgedChildChild, PreschoolClarithromycinDNA, BacterialDNA, RibosomalFemaleHumansInfantMaleMicrobial Sensitivity TestsMicrobial ViabilityMiddle AgedMycoplasma pneumoniaePharynxPneumonia, MycoplasmaPoint MutationPoint-of-Care TestingRNA, Ribosomal, 23SReagent Kits, DiagnosticReal-Time Polymerase Chain ReactionSensitivity and SpecificityYoung Adult