Reduction of radiation exposure in transcatheter atrial septal defect closure: How low must we go?

Sitefane Fidelio, Malekzadeh-Milani Sophie, Villemain Olivier, Ladouceur Magalie, Boudjemline Younes

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Cardiac catheterization relies on X-ray imaging. Most procedures are now standardized. Interventionists must strive to minimize radiation exposure to reduce the risk of induced cancers.

AIMS: To describe the radiation level in our institution, and evaluate the components contributing to radiation exposure, during transcatheter atrial septal defect (ASD) closure.

METHODS

Radiation doses for ASD closure performed between January 2009 and November 2015 were reviewed retrospectively. Data on fluoroscopic time, dose area product (DAP), DAP/kg of body weight and total air kerma were collected.

RESULTS

One hundred and seventy-four consecutive patients were included. Procedural success was 98.3%. Median procedural and fluoroscopic times were 15minutes and 1.2minutes, respectively. Median total air kerma, DAP and DAP/kg were 9.2 mGy, 88.3μGy.m2 and 3.2μGy.m2/kg, respectively. Risk factors associated with higher DAP were older age, larger ASD and device, need for balloon calibration, occurrence of complications and use of higher frame rate. Reduction of frame rate to 7.5 frames/second alone reduced by a factor of 2 the median DAP, DAP/kg and air kerma (99 vs 43μGy.m2, 3.5 vs 1.7μGy.m2/kg and 11 vs 4.8 mGy, respectively; P<0.001).

CONCLUSIONS

A low dose of radiation can be achieved for transcatheter ASD closure, even in complex ASDs, by following these recommendations: reduction of frame rate; avoidance of lateral view and cine acquisition; and limitation of fluoroscopic time by avoiding unnecessary manoeuvres and using echocardiographic guidance as much as possible.

Для цитирования: Sitefane Fidelio, Malekzadeh-Milani Sophie, Villemain Olivier, Ladouceur Magalie, Boudjemline Younes. Reduction of radiation exposure in transcatheter atrial septal defect closure: How low must we go?. Archives of cardiovascular diseases. 2017. DOI: 10.1016/j.acvd.2017.05.011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.