Медицинская библиотека · Medicina (Kaunas, Lithuania) · 2023

Ventilation Difficulty Caused by Obstructed Heated Breathing Circuit.

Kwon Jiyeon, Kang Eunsu, Shin Sunghyun, Lee Byeongcheol, Ko Myoungjin, Kim Sehun, Lee Soojee

Клинический случайСтатьяПеревод в очереди

Аннотация

For perioperative hypothermia prevention, a heated, humidified breathing circuit equipped with a fluid-warming unit inside the inspiratory limb has been developed. We report a ventilation difficulty caused by an obstructed heated breathing circuit. Cotton surrounding the hot wire, temperature sensor, and fluid tubing in the distal inspiratory limb was irregularly thicker than that of a normal circuit and nearly blocked the lumen. Despite carrying out routine checks on the anesthesia workstation preoperatively, we failed to make a prediagnosis by omitting the flow test after changing the circuit. This case puts emphasis on a routine flow test with a meticulous examination of the heated breathing circuit before every procedure.

Для цитирования: Kwon Jiyeon, Kang Eunsu, Shin Sunghyun, Lee Byeongcheol, Ko Myoungjin, Kim Sehun, Lee Soojee. Ventilation Difficulty Caused by Obstructed Heated Breathing Circuit.. Medicina (Kaunas, Lithuania). 2023. DOI: 10.3390/medicina59050965.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.