DOI 10.1007/s00247-021-05160-2
Patient-level dose monitoring in computed tomography: tracking cumulative dose from multiple multi-sequence exams with tube current modulation in children.
Аннотация
BACKGROUND
In children exposed to multiple computed tomography (CT) exams, performed with varying z-axis coverage and often with tube current modulation, it is inaccurate to add volume CT dose index (CTDIvol) and size-specific dose estimate (SSDE) to obtain cumulative dose values.
OBJECTIVE
To introduce the patient-size-specific z-axis dose profile and its dose line integral (DLI) as new dose metrics, and to use them to compare cumulative dose calculations against conventional measures.
MATERIALS AND
METHODS
In all children with 2 or more abdominal-pelvic CT scans performed from 2013 through 2019, we retrospectively recorded all series kV, z-axis tube current profile, CTDIvol, dose-length product (DLP) and calculated SSDE. We constructed dose profiles as a function of z-axis location for each series. One author identified the z-axis location of the superior mesenteric artery origin on each series obtained to align the dose profiles for construction of each patient's cumulative profile. We performed pair-wise comparisons between the peak dose of the cumulative patient dose profile and ΣSSDE, and between ΣDLI and ΣDLP.
RESULTS
We recorded dose data in 143 series obtained in 48 children, ages 0-2 years (n=15) and 8-16 years (n=33): ΣSSDE 12.7±6.7 and peak dose 15.1±8.1 mGy, ΣDLP 278±194 and ΣDLI 550±292 mGy·cm. Peak dose exceeded ΣSSDE by 20.6% (interquartile range [IQR]: 9.9-26.4%, P<0.001), and ΣDLI exceeded ΣDLP by 114% (IQR: 86.5-147.0%, P<0.001).
CONCLUSION
Our methodology represents a novel approach for evaluating radiation exposure in recurring pediatric abdominal CT examinations, both at the individual and population levels. Under a wide range of patient variables and acquisition conditions, graphic depiction of the cumulative z-axis dose profile across and beyond scan ranges, including the peak dose of the profile, provides a better tool for cumulative dose documentation than simple summations of SSDE. ΣDLI is advantageous in characterizing overall energy absorption over ΣDLP, which significantly underestimated this in all children.