Magnetic resonance in medicine · 2015
Improvement of temporal signal-to-noise ratio of GRAPPA accelerated echo planar imaging using a FLASH based calibration scan.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Talagala S Lalith, Sarlls Joelle E, Liu Siyuan, Inati Souheil J
Аннотация
PURPOSE: To demonstrate that the temporal signal-to-noise ratio (SNR) of generalized autocalibrating partially parallel acquisitions (GRAPPA) accelerated echo planar imaging (EPI) can be enhanced and made more spatially uniform by using a fast low angle shot (FLASH) based calibration scan.
METHODS: EPI of a phantom and human brains were acquired at 3 Tesla without and with GRAPPA acceleration factor of 2. The GRAPPA accelerated data were reconstructed using calibration scans acquired with EPI and FLASH acquisition schemes. The increase in temporal signal fluctuation due to GRAPPA reconstruction was quantified and compared. Simulated g-factor maps were also created for different calibration scans.
RESULTS: GRAPPA accelerated phantom data exhibited areas with high g values when using the EPI based calibration for reconstruction. The g-factor maps were uniform when using the FLASH calibration scan. g was greater than 1.1 in 74% of pixels in 64 × 64 data reconstructed with the EPI calibration compared with only 15% when using the FLASH calibration scan. Human data also showed abnormally high g regions when using the EPI calibration but not when using the FLASH calibration scan. Use of the FLASH calibration scan increased the whole brain temporal SNR by ∼12% without affecting the image quality. Experimental observations were confirmed by simulations.
CONCLUSION: A calibration scan based on a FLASH acquisition scheme can be used to improve the temporal SNR of GRAPPA accelerated EPI time series. Magn Reson Med 75:2362-2371, 2016. Published 2015. This article is a U.S. Government work and is in the public domain in the USA.
METHODS: EPI of a phantom and human brains were acquired at 3 Tesla without and with GRAPPA acceleration factor of 2. The GRAPPA accelerated data were reconstructed using calibration scans acquired with EPI and FLASH acquisition schemes. The increase in temporal signal fluctuation due to GRAPPA reconstruction was quantified and compared. Simulated g-factor maps were also created for different calibration scans.
RESULTS: GRAPPA accelerated phantom data exhibited areas with high g values when using the EPI based calibration for reconstruction. The g-factor maps were uniform when using the FLASH calibration scan. g was greater than 1.1 in 74% of pixels in 64 × 64 data reconstructed with the EPI calibration compared with only 15% when using the FLASH calibration scan. Human data also showed abnormally high g regions when using the EPI calibration but not when using the FLASH calibration scan. Use of the FLASH calibration scan increased the whole brain temporal SNR by ∼12% without affecting the image quality. Experimental observations were confirmed by simulations.
CONCLUSION: A calibration scan based on a FLASH acquisition scheme can be used to improve the temporal SNR of GRAPPA accelerated EPI time series. Magn Reson Med 75:2362-2371, 2016. Published 2015. This article is a U.S. Government work and is in the public domain in the USA.