Agreement between ccNexfin CO-trek cardiac output and intermittent cold-bolus pulmonary thermodilution in a prospective multicenter study.

Sperna Weiland Nicolaas H, de Wever Jim W, van Duivenvoorde Yoni, Boer Christa, Mitrev Ludmil, Muntazar Muhammad, Patel Kinjal, Hollmann Markus W, Preckel Benedikt

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

BACKGROUND

The ccNexfin System uses the CO-trek algorithm to analyze a non-invasively obtained arterial pressure waveform and calculate cardiac output (NEXCO). It remains matter of debate whether NEXCO can replace invasive, pulmonary artery catheter-derived, cold-bolus pulmonary thermodilution cardiac output measurement (PACCO). This study aimed at testing NEXCO-PACCO agreement in a large sample size, multicenter study. We hypothesized that agreement between NEXCO and PACCO would be demonstrated by a mean accuracy (bias) <0.6 L/min with a percentage error <30%.

METHODS

Patients undergoing cardiac surgery in three academic hospitals clinically requiring pulmonary artery catheterization were included. Exclusion criteria were aortic, pulmonary and tricuspid (valve) abnormalities, non-sinus rhythm and insufficient perfusion of the digits such as in Raynaud's disease. After induction of anesthesia, cardiac output was measured with four cold bolus thermodilution measurements and four averaged 30-second ccNexfin measurements randomized through the respiratory cycle to obtain one measurement pair. Mean accuracy and precision of ccNexfin were expressed as bias (mean of all NEXCO-PACCO differences) and limits of agreement (LOA) (1.96·SD of bias). Percentage error was calculated as [LOA / (NEXCO-PACCO average)].

RESULTS

Fifty-five patients were enrolled in the study, 51 completed the protocol. Median PACCO was 3.7 (IQR: 3.2 to 4.6) L/min and median NEXCO was 3.8 (IQR: 3.1 to 4.7) L/min. NEXCO-PACCO bias was 0.1 (LOA: -1.4 to +1.6) L/min with a 37% percentage error.

CONCLUSIONS

In this study, cardiac output measurement with ccNexfin failed to meet the predefined criteria for agreement with cold-bolus pulmonary artery thermodilution.

Для цитирования: Sperna Weiland Nicolaas H, de Wever Jim W, van Duivenvoorde Yoni, Boer Christa, Mitrev Ludmil, Muntazar Muhammad, Patel Kinjal, Hollmann Markus W, Preckel Benedikt. Agreement between ccNexfin CO-trek cardiac output and intermittent cold-bolus pulmonary thermodilution in a prospective multicenter study.. Minerva anestesiologica. 2017. DOI: 10.23736/s0375-9393.17.12051-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.