Infective endocarditis after device closure of atrial septal defects: Case report and review of the literature.
Аннотация
We report a case of late infective endocarditis in an 8-year-old boy 3 years after transcatheter closure of an atrial septal defect with the Amplatzer Atrial Septal Occluder Device. Echocardiography showed a very thick pannus lining the left atrial disc of the prosthesis, with some mobile elements. MRI showed several cerebral microemboli. The patient had dental caries and blood cultures found Staphylococcus aureus. Patient was treated with gentamicin and oxacillin before surgical removal of the prosthesis. Twenty-one cases of infective endocarditis after atrial defect device closure have been reported in the literature (13 ostium secundum ASD and 8 patent foramen ovale). Seven pediatric cases were reported. S. aureus was the most frequent bacterium. Antibiotics were administered in all cases and most patients (n = 18) underwent surgical removal of the device. Incomplete endothelialization of the prosthesis was a suggested mechanism. Two patients died after surgery. © 2016 Wiley Periodicals, Inc.