DOI 10.1016/j.acvd.2015.12.003
The new Occlutech(®) patent ductus arteriosus occluder: Single centre experience.
Аннотация
BACKGROUND
Transcatheter therapy is the first-line treatment for closure of patent ductus arteriosus (PDA). A range of different devices have been used.
AIM: To report our experience using the new Occlutech(®) PDA occluder for closure of PDA.
METHODS
All consecutive patients receiving an Occlutech(®) PDA occluder for closure of PDA from March 2014 to May 2015 were included in the study. Occluders were implanted using a conventional technique. After cardiac catheterization, transthoracic echocardiography was performed sequentially. Residual shunting and procedural and mid-term complications, including embolization, device-induced left pulmonary artery stenosis and aortic coarctation, were systematically assessed and reported.
RESULTS
Fifty-six patients (35 females) with a median age of 1 year (range: 29 days to 24 years) were included in the study. PDAs were closed successfully using device number 5/3.5 in 17 patients, 6/4 in 28 patients, 7/5 in four patients, 8/6 in six patients and 10/8 in one patient. Post-implantation angiography showed no residual shunt in 10 patients (17.8%), moderate intraprosthetic shunt in 43 patients (76.8%) and severe shunt in three patients (5.4%). One device embolized in the right pulmonary artery 1hour after successful closure; the device was retrieved surgically after unsuccessful snaring. After a mean follow-up of 10 months (range: 3-17.8 months), no residual shunt or device-related complications were noted.
CONCLUSION
The new Occlutech(®) PDA occluder allows safe and efficient PDA closure, similar to existing devices.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.