MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.27974

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2018

Implantation of the Melody transcatheter pulmonary valve PB1016 in patients with dysfunctional right ventricular outflow tract conduits.

Morray Brian H, Jones Thomas K, Coe James Y, Gitter Roland, Martinez Jose Zunzunegui, Turner Daniel R, Gray Robert G, Lung Te-Hsin, Berman Darren P, Levi Daniel S

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

OBJECTIVES

This study describes procedural and 1-year outcomes of the 16 mm Melody PB1016 valve in patients with dysfunctional RVOT conduits.

BACKGROUND

The Melody PB1016 is a standard Melody valve produced from a 16 mm bovine jugular vein and is intended for deployment up to 20 mm.

METHODS

This is a prospective, non-randomized, multicenter study of the procedural and short-term outcomes of Melody PB1016 TPV replacement within dysfunctional RVOT conduits. Data from eight centers were included in the analysis.

RESULTS

During the study period, 39 patients underwent attempted Melody TPVR. Of the 39 patients, 30 underwent successful Melody TPVR. The majority of patients underwent placement of one or more stents prior to TPVR. There was a significant reduction in peak conduit pressure gradient following TPVR (38 mmHg vs. 11 mmHg, P < 0.001). There were three cases of confined conduit tears successfully treated with covered stents or the valve itself. Repeat catheterization was performed in one patient for early re-obstruction that was successfully treated with balloon valvuloplasty. At recent follow-up, there were no cases of more than mild valve regurgitation and the mean pulmonary valve gradient by echocardiogram remained reduced relative to pre-TPVR implant measurements (33.5 mmHg vs. 15.2 mmHg). There were no cases of valve stent fracture or endocarditis reported at the 1-year follow-up.

CONCLUSIONS

Our analysis of TPVR with the PB1016 valve in RVOT conduits showed it to be safe and effective and can be performed in a wide range of conduit sizes with preserved valve function. ClinicalTrials.gov Identifier: NCT02347189.

Для цитирования: Morray Brian H, Jones Thomas K, Coe James Y, Gitter Roland, Martinez Jose Zunzunegui, Turner Daniel R, Gray Robert G, Lung Te-Hsin, Berman Darren P, Levi Daniel S. Implantation of the Melody transcatheter pulmonary valve PB1016 in patients with dysfunctional right ventricular outflow tract conduits.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2018. DOI: 10.1002/ccd.27974.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.