[Improvement of registration accuracy for navigated-control drill in mastoidectomy (autopilot)].

Strauss G, Dittrich E, Baumberger C, Hofer M, Strauss M, Stopp S, Koulechov K, Dietz A, Lüth T

Аннотация на английскомСтатьяПеревод в очереди

Аннотация

BACKGROUND

The goal of this study is the improvement of the surgical accuracy of a navigate-controlled drill for mastoidectomy in a lab test.

METHODS

For lab tests an artificial model of the temporal bone with color-coded injury identification of the facial nerve (solution of 0.5 mm) was used. Two different registration methods were examined: (group 1) navigation bow with 4 integrated markers at the upper jaw; (group 2) landmark registration with 4 titanium micro screws. An optical navigation system was used. The targets were illustrated by 3 titanium screws within the range of the planum mastoideum. The accuracy of the navigate-controlled drill in drilling the planned cavity were evaluated at 20 temporal bone models. The measurement of the registration accuracy was evaluated by deviation between the target screw and the calculated position in the navigation system. The evaluation of the resulted cavities was done by 5 senior surgeons with the help of the microscope.

RESULTS

The registration accuracy shows a maximum deviation between the real position and the calculated position of 1,73 MM in group of 1 and 0.93 MM in group 2. In group 1 the nerve was hurt in 5/20 cases and a maximum deviation of - 1.5 mm (Std 0.25 mm) (drilled beyond the nerve) was measured. In group 2 the nerve was not hurt, a maximum deviation of 0.5 mm (too early stopped before the nerve) was measured.

CONCLUSIONS

Significantly better results of the registration and drilling accuracy show up in group group 2. Thus the preconditions for clinical use are fulfilled.

Для цитирования: Strauss G, Dittrich E, Baumberger C, Hofer M, Strauss M, Stopp S, Koulechov K, Dietz A, Lüth T. [Improvement of registration accuracy for navigated-control drill in mastoidectomy (autopilot)].. Laryngo- rhino- otologie. 2008. DOI: 10.1055/s-2007-995629.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.