DOI 10.1259/dmfr/28020783

Медицинская библиотека · Dento maxillo facial radiology · 2005

Accuracy of measurements of mandibular anatomy and prediction of asymmetry in panoramic radiographic images.

Laster W S, Ludlow J B, Bailey L J, Hershey H Garland

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVES

Measurements of ideally positioned and systematically mis-positioned skulls were used to evaluate errors in linear measurements and symmetry ratios made with panoramic X-ray images.

METHODS

Digital panoramic images of 30 skulls placed in ideal, shifted and rotated positions, were assessed by measuring distances between anatomic points and fiducial references. Differences between photographic measurements (control) and radiographic measurements were compared. Horizontal measurements included a 20 mm wire and the distance from gonion to mental foramen (G-MF). Vertical distances measured included a 40 mm wire, condyle to sigmoid notch length, and condyle to gonion (posterior mandibular height or PMH). A relative symmetry ratio comparing the difference between right and left PMH was also calculated. Distances measured in panoramic images were corrected using the left vertical wire distance or the panoramic unit's stated magnification factor (1.25x).

RESULTS

Greatest differences were noted for horizontal measurements and shifted skull positions. Use of an arbitrary magnification correction was consistently less accurate than use of an internal calibration and resulted in general underestimation of actual dimensions. Measures of PMH varied significantly from expected values for each of the three skull positions (P<0.005). Panoramic accuracy for detecting asymmetry was 67% for ideal, 70% for rotated, and 47% for shifted skull positions when an internal reference was used.

CONCLUSIONS

Panoramic radiographs should be used with caution in making absolute measurements or relative comparisons. Even when internal fiducial calibration for image distortion of anatomy is used, measurements such as those assessing posterior mandibular facial symmetry may be unreliable.

Для цитирования: Laster W S, Ludlow J B, Bailey L J, Hershey H Garland. Accuracy of measurements of mandibular anatomy and prediction of asymmetry in panoramic radiographic images.. Dento maxillo facial radiology. 2005. DOI: 10.1259/dmfr/28020783.