Patient Cranial Angle and Intrafractional Stability in CyberKnife Robotic Radiosurgery: A Retrospective Analysis.

Kang Chen-Lin, Huang Ya-Yu, Chen Yi-Ren, Tsai Shu-Huei, Huang Chun-Chieh, Huang Yu-Jie

СтатьяПеревод в очереди

Аннотация

Purpose: The aim of this study was to investigate whether variations in cranial angles and treatment accuracy during CyberKnife robotic radiosurgery necessitate adjustment of the margins of the planning target volume. Patients and

Methods

Data from 66 patients receiving CyberKnife treatment for brain tumors were retrospectively analyzed. Patients were immobilized using a thermoplastic mask and headrest. The cranial angle was measured on planning CT and patients were divided into 2 groups: ≤10° (Group A) and >10° (Group B). Intrafractional motion was recorded using the CyberKnife tracking system over 50 min. Translational and rotational errors were compared between groups, and planning target volume margins were calculated.

Results

In Group A, significant translational error differences were found along with the X-axis over time (P < .02). In Group B, significant differences occurred along with the Z-axis (P < .03). No significant rotational or 3-dimensional vector differences were found in either group. Group A had significantly lower Y-axis (P < .045) and roll axis (P < .005) errors compared to Group B. Estimated planning target volume margins in Group A were 0.56 mm (X), 0.46 mm (Y), and 0.47 mm (Z). In Group B, margins were 0.62 mm (X), 0.48 mm (Y), and 0.46 mm (Z). Margins covering 95% of intrafraction motion were 0.49 to 0.50 mm (X, Y, Z) and 0.69 mm (3-dimensional vector) for Group A, and 0.48 to 0.60 mm and 0.79 mm for Group B. With a 1-mm margin, complete coverage was achieved in Group A while 2.1% of vectors in Group B exceeded 1 mm.

Conclusion

Adjusting cranial angle to ≤10° during thermoplastic mask molding provided better or similar intrafractional stability compared to >10°.

Для цитирования: Kang Chen-Lin, Huang Ya-Yu, Chen Yi-Ren, Tsai Shu-Huei, Huang Chun-Chieh, Huang Yu-Jie. Patient Cranial Angle and Intrafractional Stability in CyberKnife Robotic Radiosurgery: A Retrospective Analysis.. Technology in cancer research & treatment. 2024. DOI: 10.1177/15330338241235058.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.