MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1515/jpm.2003.012

Медицинская библиотека · Journal of perinatal medicine · 2003

New transcutaneous jaundice device with two optical paths.

Yasuda Saneyuki, Itoh Susumu, Isobe Kenichi, Yonetani Masahiko, Nakamura Hajime, Nakamura Makoto, Yamauchi Yoshitada, Yamanishi Akio

Клиническое исследованиеСравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVES

To confirm the accuracy and precision of transcutaneous bilirubin (TcB) values measured by a new device with two optical paths (JM-103) and the value of total serum bilirubin (TSB) level in clinical units of measurement.

METHODS

For comparison of the levels of accuracy and precision of JM-103 and the old device (JM-102), serum samples were collected from 77 Japanese infants in three different hospitals including 24 preterm infants and 53 term infants. Measurement of TcB by JM-103 and JM-102 were performed on the forehead of each infant within 30 min before or after blood sampling.

RESULTS

The range of TSB was limited to 19.6 mg/dL and to 17 mg/dL for preterm infants. The correlation coefficients for all subjects (r = 0.94) and for term and preterm subjects between TcB measured by JM-103 and TSB was higher than that between TcB measured by JM-102 and TSB. The regression line in term infants between TcB measured by JM-103 and TcB was similar to that in preterm infants. The error distribution of TcB measured by JM-102 and TSB for all subjects (0.00 +/- 2.21) and for term and preterm subjects was larger than that of TcB measured by JM-103 and TSB (all subjects, 0.30 +/- 1.55).

Для цитирования: Yasuda Saneyuki, Itoh Susumu, Isobe Kenichi, Yonetani Masahiko, Nakamura Hajime, Nakamura Makoto, Yamauchi Yoshitada, Yamanishi Akio. New transcutaneous jaundice device with two optical paths.. Journal of perinatal medicine. 2003. DOI: 10.1515/jpm.2003.012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.