DOI 10.1016/j.annemergmed.2015.07.011
Lacerations and Embedded Needles Caused by Epinephrine Autoinjector Use in Children.
Аннотация
STUDY
OBJECTIVE
Epinephrine autoinjector use for anaphylaxis is increasing. There are reports of digit injections because of incorrect autoinjector use, but no previous reports of lacerations, to our knowledge. We report complications of epinephrine autoinjector use in children and discuss features of these devices, and their instructions for use, and how these may contribute to injuries.
METHODS
We queried emergency medicine e-mail discussion lists and social media allergy groups to identify epinephrine autoinjector injuries involving children.
RESULTS
Twenty-two cases of epinephrine autoinjector-related injuries are described. Twenty-one occurred during intentional use for the child's allergic reaction. Seventeen children experienced lacerations. In 4 cases, the needle stuck in the child's limb. In 1 case, the device lacerated a nurse's finger. The device associated with the injury was operated by health care providers (6 cases), the patient's parent (12 cases, including 2 nurses), educators (3 cases), and the patient (1 case). Of the 3 epinephrine autoinjectors currently available in North America, none include instructions to immobilize the child's leg. Only 1 has a needle that self-retracts; the others have needles that remain in the thigh during the 10 seconds that the user is instructed to hold the device against the leg. Instructions do not caution against reinjection if the needle is dislodged during these 10 seconds.
CONCLUSION
Epinephrine autoinjectors are lifesaving devices in the management of anaphylaxis. However, some have caused lacerations and other injuries in children. Minimizing needle injection time, improving device design, and providing instructions to immobilize the leg before use may decrease the risk of these injuries.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.