Evaluation of Noninvasive Hemoglobin Monitoring in Trauma Patients with Low Hemoglobin Levels.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Gamal Medhat, Abdelhamid Bassant, Zakaria Dina, Dayem Omnia Abd El, Rady Ashraf, Fawzy Maher, Hasanin Ahmed

Аннотация

OBJECTIVE: Bleeding is a leading cause of death among trauma patients. Delayed assessment of blood hemoglobin level might result in either unnecessary blood transfusion in nonindicated patients or delayed blood transfusion in critically bleeding patients. In this study, we evaluate the precision of noninvasive hemoglobin monitoring in trauma patients with low hemoglobin levels.

METHODS: We included trauma patients with low hemoglobin levels (less than 8 g/dL) scheduled for surgical intervention. Blood samples were obtained on admission and after each blood unit with concomitant measurement of serum hemoglobin using radical-7 Masimo device. The change in blood hemoglobin after every transfused blood unit was also assessed by both methods (change in noninvasive Masimo hemoglobin [Delta-Sp-Hb] and change in laboratory hemoglobin [Delta-Lab-Hb]). The precision of Masimo hemoglobin level (Sp-Hb) compared with Laboratory hemoglobin level (Lab-Hb) was determined using both Bland-Altman and Pearson correlation analyses.

RESULTS: One hundred eighty-four time-matched samples were available for final analysis. Bland-Altman analysis showed excellent accuracy of Sp-Hb compared with Lab-Hb with mean bias of 0.12 g/dL and limits of agreement between -0.56 g/dL and 0.79 g/dL. Excellent correlation was reported between both measures with Pearson correlation coefficient of 0.872. Excellent agreement was also reported between both Delta-Sp-Hb and Delta-Lab-Hb with mean bias of -0.05 and limits of agreement from -0.62 to 0.51 CONCLUSIONS:: Sp-Hb showed accurate precision in both absolute values and trend values compared with Lab-Hb measurement in trauma patients with low hemoglobin levels.