DOI 10.1016/j.arth.2021.01.086
High Rates of Spacer Fracture in the Setting of Extended Trochanteric Osteotomy With a Specific Thin-Core Articulating Antibiotic Hip Spacer.
Аннотация
BACKGROUND
Two-stage revision remains the standard of care for prosthetic joint infection after total hip arthroplasty. However, there are substantial complications associated with articulating antibiotic hip spacers. Handmade and molded spacers have been shown to have higher rates of spacer fracture than antibiotic-coated prostheses (ACPs). The aim of this study is to review outcomes with an implant that is often categorized as an ACP spacer, the Zimmer-Biomet StageOne Select Femoral Spacer (ZBSO).
METHODS
A retrospective review was performed of 63 patients who underwent placement of a ZBSO. Patients were compared based on whether or not an extended trochanteric osteotomy (ETO) was performed using Fisher's exact and t-tests.
RESULTS
Five patients were excluded due to lack of follow-up or death shortly after stage 1 surgery, leaving 58 patients. Spacer fracture was noted in 5 of 58 patients (8.6%). Sixteen patients underwent ETO and 25.0% suffered a spacer fracture compared to 2.3% without ETO (odds ratio 13.7, P = .0248). There was no association between patient demographics or ETO length and spacer fracture. Two patients had periprosthetic fractures (3.4%) and 4 had dislocations (6.9%). Forty-nine patients (84.4%) went on to second-stage revision; of those 26.5% failed to clear the infection and required an average of 2.2 additional surgeries.
CONCLUSION
The ZBSO spacer has overall complication rates similar to previously reported spacer series. Although the ZBSO looks like an ACP spacer, in the setting of ETO, it behaves like a molded or handmade spacer with a high rate of spacer fracture (25%) due to the small diameter of the core. This implant should be used with caution in combination with an ETO.