Safety and behavior of implantable electronic devices during cremation.

Maher Timothy, Kong Nathan, Spray Ryan, Lee Sophie, Gurgel Sofia, Waks Jonathan, Kramer Daniel B, Ellenbogen Kenneth A, Zimetbaum Peter, d'Avila Andre

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The current standard of practice for cremating patients with cardiac implantable electronic devices (CIEDs) is surgical explantation before cremation to mitigate the risk of device explosion. This surgery may conflict with patient or family beliefs, whereas cremation of CIEDs may create occupational hazards.

OBJECTIVE

This study sought to establish an ex vivo model for screening CIED behavior during cremation.

METHODS

Seven CIEDs underwent testing including projectile/sound testing, impact testing, and gas analysis. In the projectile test, devices were heated until thermal failure (explosion) and filmed with a high-speed camera and microphone. For impact testing, brick structures were built to assess damage after explosion. Gas chromatography-mass spectrometry identified released gases. Findings were compared with occupational health standards, where available.

RESULTS

The implantable loop recorder and leadless pacemaker produced minimal kinetic energy and impact risk with thermal failure. The remaining devices demonstrated explosive disintegration at thermal temperatures <500°C. The pacemakers and implantable cardiac defibrillators produced sound levels >120 dB and resulted in damage to brick structures. Small quantities of benzene and hydrogen fluoride were produced but at quantities within acceptable occupational exposure limits in a cremation chamber.

CONCLUSION

All tested CIEDs experienced explosion at temperatures below crematorium standards. The smallest devices produced minimal risk of damage or injury, suggesting that they may safely remain in situ during cremation, whereas the larger devices produced more kinetic energy, testing chamber damage, and louder explosions, suggesting potential risk with cremation. Cadaveric testing in full-sized cremation chambers is required to determine real-world risk.

Для цитирования: Maher Timothy, Kong Nathan, Spray Ryan, Lee Sophie, Gurgel Sofia, Waks Jonathan, Kramer Daniel B, Ellenbogen Kenneth A, Zimetbaum Peter, d'Avila Andre. Safety and behavior of implantable electronic devices during cremation.. Heart rhythm. 2024. DOI: 10.1016/j.hrthm.2024.08.031.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.