Novel methods for the quantification of toxic, residual phase transfer catalyst in fluorine-18 labeled radiotracers.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Blevins David W, Rigney Grant H, Fang Michael Y, Akula Murthy R, Osborne Dustin R

Аннотация

INTRODUCTION: Fluorine-18 labeled radiopharmaceuticals undergo quality control testing for residual phase-transfer-catalyst content. The almost universally used quality-control test is a silica plate spot-test comparison of the radiopharmaceutical beside a 50-ppm standard. Once developed by staining, the radiopharmaceutical spot must be of equal or less intensity to pass the test. There is currently a need for a quantitative, inexpensive, and less subjective quality control method that allows the automatic incorporation of the acquired measurement directly into electronic batch reports.

RESULTS: In the developed method, a resazurin test solution is mixed with an aliquot of the radiopharmaceutical analyte along with dichloromethane (DCM). The mixture is vortexed. The potassium resazurin-phase transfer catalyst complex solubilizes into the DCM imparting a blue color. The organic layer is then removed for analysis. Three measurement methods were utilized: visual colorimetry against pre-prepared standards, spectrophotometric measurement of transmittance, and electrical conductance. A simple prototype spectrophotometer and an electrical test cell were constructed to acquire data. Sodium Resazurin dye was found to be a suitable test chromophore for residual phase transfer catalyst analysis of aqueous solutions. Quantitative spectrophotometric measurements are possible in the 0-100-ppm range (18-crown-6) and 0-150-ppm range (Kryptofix® or tetrabutylammonium). Electrical resistance measurements of the phase transfer-catalyst resazurin complex in DCM are also a viable method, allowing quantitative phase transfer catalyst measurements in the 0-100-ppm range.

CONCLUSION: The methodologies developed are more quantitative alternatives to the current spot-test method. The spectrophotometric method was determined to be the most accurate method.