Медицинская библиотека · Journal of cancer research and therapeutics · 2017

Evaluation of the effect of temperature variation on response of PRESAGE® dosimeter.

Farhood Bagher, Khezerloo Davood, Zadeh Tohid Morteza, Nedaie Hassan Ali, Hamrahi Daryoosh, Khezerloo Najimeh

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Many factors, such as PRESAGE ® composition, dose rate, energy, and type of radiation, temperature, etc., may effect on PRESAGE ® dosimeter response. The aim of this study was investigating the effect of temperature variation on response of PRESAGE ® solid dosimeter.

MATERIALS AND

METHODS

In this study, a PRESAGE ® solid detector was fabricated. Ninety-four percent weight polyurethane, 5% weight carbon tetrachloride, and 1% weight leucomalachite green were used. Radiological and physical characteristics of PRESAGEs ®, such as mass density, electron density, and effective number atomic were obtained and compared with water. Response of PRESAGE ® dosimeter in temperatures -4, 10, 25, 35, 45, 55, 65, 75, 85, and 90°C was evaluated. In addition, the absorption peak at various temperatures was investigated.

RESULTS

The results showed that the absorption peak at different temperatures was in the range of 630-635 nm. For temperatures below 75°C, the results indicated that temperature variation has no effect on the response of PRESAGE ® dosimeter whereas at the temperatures> 75°C, temperature variation has an effect on PRESAGE ® dosimeter response.

CONCLUSION

The finding showed that temperature changes have not impact on the absorption peak. In addition, the results related to the effect of temperature variation on the response of PRESAGE ® dosimeter showed that in the range of clinical applications (temperatures below 75°C), temperature variation has no effect on PRESAGE ® dosimeter response.

Для цитирования: Farhood Bagher, Khezerloo Davood, Zadeh Tohid Morteza, Nedaie Hassan Ali, Hamrahi Daryoosh, Khezerloo Najimeh. Evaluation of the effect of temperature variation on response of PRESAGE® dosimeter.. Journal of cancer research and therapeutics. 2017. DOI: 10.4103/0973-1482.199389.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.