DOI 10.1016/j.ajem.2015.11.001
High-frequency linear transducer improves detection of an intrauterine pregnancy in first-trimester ultrasonography.
Аннотация
OBJECTIVE
The objective was to determine if the need for transvaginal ultrasonographic examination can be decreased by the addition of the transabdominal high-frequency, 12-4-MHz linear transducer after a failed examination with the 6-2-mHz curvilinear transducer when evaluating for an intrauterine pregnancy (IUP).
METHODS
This is a prospective pilot study of women in their first trimester of pregnancy presenting to the emergency department with abdominal pain and/or vaginal bleeding. If no IUP was identified using the curvilinear transducer via the transabdominal approach, they were subsequently scanned using the linear transducer. Patients without evidence of an IUP transabdominally were scanned via the transvaginal approach.
RESULTS
Eighty-one patients were evaluated; no IUP was visualized in 27 using the standard curvilinear transducer approach, and these then had an ultrasonography performed with the linear transducer. Of these, 9 patients (33.3%; 0.95 confidence interval [CO], 15.5%-51.1%) were found to have an IUP with the linear transducer. For the 18 patients who received a transvaginal scan, 15 patients (83.3%; 0.95 CI, 66.1%-100%) had no IUP identified with the transvaginal transducer, and 3 (16.7%; 0.95 CI, 0%-33.9%) had an IUP identified.
CONCLUSIONS
The transabdominal use of a high-frequency linear transducer in the evaluation of patients in the first trimester after failed curvilinear transducer results in a clinically significant reduction in the need for transvaginal ultrasonography to confirm the presence of an IUP.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.