DOI 10.5664/jcsm.2148

Медицинская библиотека · Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine · 2012

Detection of sleep disordered breathing and its central/obstructive character using nasal cannula and finger pulse oximeter.

Sommermeyer Dirk, Zou Ding, Grote Ludger, Hedner Jan

Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

STUDY

OBJECTIVE

To assess the accuracy of novel algorithms using an oximeter-based finger plethysmographic signal in combination with a nasal cannula for the detection and differentiation of central and obstructive apneas. The validity of single pulse oximetry to detect respiratory disturbance events was also studied.

METHODS

Patients recruited from four sleep laboratories underwent an ambulatory overnight cardiorespiratory polygraphy recording. The nasal flow and photoplethysmographic signals of the recording were analyzed by automated algorithms. The apnea hypopnea index (AHI(auto)) was calculated using both signals, and a respiratory disturbance index (RDI(auto)) was calculated from photoplethysmography alone. Apnea events were classified into obstructive and central types using the oximeter derived pulse wave signal and compared with manual scoring.

RESULTS

Sixty-six subjects (42 males, age 54 ± 14 yrs, body mass index 28.5 ± 5.9 kg/m(2)) were included in the analysis. AHI(manual) (19.4 ± 18.5 events/h) correlated highly significantly with AHI(auto) (19.9 ± 16.5 events/h) and RDI(auto) (20.4 ± 17.2 events/h); the correlation coefficients were r = 0.94 and 0.95, respectively (p < 0.001) with a mean difference of -0.5 ± 6.6 and -1.0 ± 6.1 events/h. The automatic analysis of AHI(auto) and RDI(auto) detected sleep apnea (cutoff AHI(manual) ≥ 15 events/h) with a sensitivity/specificity of 0.90/0.97 and 0.86/0.94, respectively. The automated obstructive/central apnea indices correlated closely with manually scoring (r = 0.87 and 0.95, p < 0.001) with mean difference of -4.3 ± 7.9 and 0.3 ± 1.5 events/h, respectively.

CONCLUSIONS

Automatic analysis based on routine pulse oximetry alone may be used to detect sleep disordered breathing with accuracy. In addition, the combination of photoplethysmographic signals with a nasal flow signal provides an accurate distinction between obstructive and central apneic events during sleep.

Для цитирования: Sommermeyer Dirk, Zou Ding, Grote Ludger, Hedner Jan. Detection of sleep disordered breathing and its central/obstructive character using nasal cannula and finger pulse oximeter.. Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine. 2012. DOI: 10.5664/jcsm.2148.

Предметные рубрики MeSH