A comparison between size of the occluder device and two-dimensional transoesophageal echocardiographic sizing of the ostium secundum atrial septal defect.

Hajizeinali Alimohammad, Sadeghian Hakimeh, Rezvanfard Mehrnaz, Alidoosti Mohammad, Zoroufian Arezoo, Volman Marat A

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVES

Transcatheter closure of a secundum atrial septal defect (ASD II) has become an effective alternative for surgical treatment. In this study we evaluated the correlation between the two-dimensional transoesophageal echocardiographic (2D TEE) sizing of ASDs and the actual diameter of occluders in patients undergoing device closure.

METHODS

The records of 54 patients who underwent transcatheter ASD closure were reviewed. ASD characteristics and maximum defect diameter were evaluated using preprocedure 2D TEE images. Appropriate device size was determined by the balloon sizing method, which measures the balloon occlusive diameter (BOD) via TEE and fluoroscopy. ASD closure was performed under continuous TEE monitoring using the Amplatzer occluder in all patients.

RESULTS

The mean of the TEE-derived maximum defect diameter was significantly lower than the mean of the BOD (17.8 ± 4.5 vs 22.1 ± 5.1 mm; p < 0.001) and the mean size of the implanted occluder device (17.8 ± 4.5 vs 23.3 ± 5.1 mm; p < 0.001). However, a good correlation was found between the TEE-derived defect size and the BOD (BOD = 0.898 × TEE defect size + 6.212, R = 0.824; p < 0.001) and between the TEE measurement and the final size of the implanted Amplatzer (device size = 0.928 × TEE defect size + 6.853, R = 0.822; p < 0.001).

CONCLUSION

2D TEE may provide a good equation to predict the BOD or the size of the occluder device; however, further studies are needed to investigate whether it is feasible to perform transcatheter ASD occlusion without balloon sizing.

Для цитирования: Hajizeinali Alimohammad, Sadeghian Hakimeh, Rezvanfard Mehrnaz, Alidoosti Mohammad, Zoroufian Arezoo, Volman Marat A. A comparison between size of the occluder device and two-dimensional transoesophageal echocardiographic sizing of the ostium secundum atrial septal defect.. Cardiovascular journal of Africa. 2013. DOI: 10.5830/cvja-2013-014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.