MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1212/wnl.0b013e318228c0fb

Медицинская библиотека · Neurology · 2011

Patient-ventilator asynchrony with nocturnal noninvasive ventilation in ALS.

Atkeson A D, RoyChoudhury A, Harrington-Moroney G, Shah B, Mitsumoto H, Basner R C

Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

American Academy of Neurology (AAN) practice parameters for amyotrophic lateral sclerosis (ALS) include noninvasive positive pressure ventilation (NIV) for a forced vital capacity (FVC) ≤50%predicted. Despite the limited ability of NIV systems to deliver adequate ventilation synchronous with patient demand, nocturnal patient-ventilator asynchrony has not been systematically studied in patients with ALS prescribed such NIV.

METHODS

Twenty-three consecutively recruited patients with ALS reporting consistent use of nocturnal NIV (nNIV) prescribed for FVC ≤50% or orthopnea underwent home nocturnal polysomnography (PSG) on their current nNIV regimen (all used bilevel positive airway pressure). PSG recorded airflow, NIV pressures, thoracic and abdominal respiratory effort, and O(2) saturation by pulse oximetry (SpO(2)). Patient-ventilator asynchrony was calculated as an asynchrony index (AI), the number of episodes of asynchrony per hour of nocturnal recording time (RT).

RESULTS

Nineteen patients had an adequate PSG. Their mean AI was 69 ± 46 SD (range 15-146/hour). Mean asynchrony time as a percent of RT was 17% ± 19%. Mean nadir SpO(2) was 85% ± 7%. In multiple regression analysis, no demographic, functional severity (including FVC and ALS Revised Functional Rating Scale), or NIV (including pressure levels and duration of NIV prescription) variables were significantly predictive of degree of patient-ventilator asynchrony.

CONCLUSIONS

These findings document frequent nocturnal patient-ventilator asynchrony in patients with ALS consistently using nNIV prescribed as per current AAN practice parameters, and suggest that use of nNIV per these parameters is unlikely to provide patients with ALS optimal nocturnal ventilatory support.

Для цитирования: Atkeson A D, RoyChoudhury A, Harrington-Moroney G, Shah B, Mitsumoto H, Basner R C. Patient-ventilator asynchrony with nocturnal noninvasive ventilation in ALS.. Neurology. 2011. DOI: 10.1212/wnl.0b013e318228c0fb.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.