Fluoroscopically-guided manipulation of malfunctioning peritoneal dialysis catheters.
Аннотация
OBJECTIVE
To review our experience with fluoroscopic evaluation and manipulation of malpositioned, malfunctioning, peritoneal dialysis (PD) catheters.
MATERIALS AND
METHODS
Thirty-one patients, over a 5-year period (1 May 1992 to 30 April 1997) with malfunctioning PD catheters, who had fluoroscopically-guided manipulation were reviewed. Catheters were manipulated using a malleable aluminum bar and, if necessary, guide wires or other stiffeners. Technical success was assessed on the basis of adequate, fluoroscopically verified, catheter placement at the time of the procedure and improved flows. A functional PD catheter at 30 days post manipulation was considered to be a clinically successful manipulation.
RESULTS
There were 41 manipulations [33 initial (IM) and 8 remanipulations (RM)] for malpositioned or kinked catheters. In 31 (19 male, 12 female) patients ranging in age from 31 to 76 years (mean age 60 years), the initial technical success rate was 85% for IM (n = 28/33) and 63% (n = 5/8) for RM. The overall clinical success rate, or 30-day primary patency, was 55% for IM (n = 18/33) and 63% for RM. Catheter function (combined IM and RM) continued for a median 869 days (95% CI: 118, 1620). No early complications were noted.
CONCLUSIONS
Fluoroscopic manipulation, including attempts at remanipulation, of PD catheters is a safe procedure. The technique is a simple, inexpensive, and effective way of prolonging PD catheter life, thereby reducing the number of surgical interventions.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.