Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2011
Transcatheter closure of large patent ductus arteriosus at high altitude with a novel nitinol device.
СтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Heath Alexandra, Lang Nora, Levi Daniel S, Granja Miguel, Villanueva Jorge, Navarro Joaquin, Echazú Gabriel, Kozlik-Feldmann Rainer, del Nido Pedro, Freudenthal Franz
Аннотация
BACKGROUND: Patent ducti arteriosi (PDAs) are more frequent and larger at high altitude than at sea level. A novel PDA closure device, the Nitocclud PDA-R, is designed specifically for both large and medium size PDAs. The initial clinical experience with a new nitinol-based device in high altitude patients with large PDAs is described.
METHODS: The Nitocclud PDA-R is a self-expandable, self-centering, repositionable occluder made of one nitinol wire without use of welding. It contains several polyester membranes, is delivered with a central guide wire and is released by retraction of the central wire into the delivery catheter. The efficacy of this device was evaluated at several high altitude centers.
RESULTS: Fifty-one patients without other congenital cardiac defects underwent transcatheter closure of PDA. Complete occlusion of the PDA was achieved in 98% of the patients. Nearly 49% of the patients had no shunt immediately after device implantation. Echocardiography revealed a complete closure rate by Doppler interrogation of 69% after 24 hr, 96% after 6 months, and 98% after 1 year. In two cases, device embolization was observed after release, and in both cases the device was easily retrieved with standard interventional techniques. There have been no episodes of delayed device migration, endocarditis, hemolysis, wire fracture, device disruption, or death.
CONCLUSIONS: The Nitocclud PDA-R device is safe and effective and can easily close very large PDAs. This device has a high rate of complete occlusion within 1 year and is easily retrieved if embolized.
METHODS: The Nitocclud PDA-R is a self-expandable, self-centering, repositionable occluder made of one nitinol wire without use of welding. It contains several polyester membranes, is delivered with a central guide wire and is released by retraction of the central wire into the delivery catheter. The efficacy of this device was evaluated at several high altitude centers.
RESULTS: Fifty-one patients without other congenital cardiac defects underwent transcatheter closure of PDA. Complete occlusion of the PDA was achieved in 98% of the patients. Nearly 49% of the patients had no shunt immediately after device implantation. Echocardiography revealed a complete closure rate by Doppler interrogation of 69% after 24 hr, 96% after 6 months, and 98% after 1 year. In two cases, device embolization was observed after release, and in both cases the device was easily retrieved with standard interventional techniques. There have been no episodes of delayed device migration, endocarditis, hemolysis, wire fracture, device disruption, or death.
CONCLUSIONS: The Nitocclud PDA-R device is safe and effective and can easily close very large PDAs. This device has a high rate of complete occlusion within 1 year and is easily retrieved if embolized.
Предметные рубрики MeSH
AdolescentAdultAlloysAltitudeArgentinaBoliviaCardiac CatheterizationChildChild, PreschoolDevice RemovalDuctus Arteriosus, PatentEchocardiography, Doppler, ColorEquipment DesignFemaleForeign-Body MigrationHemodynamicsHumansInfantMaleMembranes, ArtificialPolyestersProspective StudiesTherapeuticsTime FactorsYoung Adult