Effectiveness of Platelet Function Analyzer-100 for Laboratory Detection of Anti-Platelet Drug-Induced Platelet Dysfunction.

Kweon Oh Joo, Lim Yong Kwan, Kim Bohyun, Lee Mi Kyung, Kim Hye Ryoun

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

High on-treatment platelet reactivity (HTPR) is the phenomenon wherein patients exhibit normal platelet activity in laboratory testing despite adequate adherence to anti-platelet treatment. We investigated the detection rates of Platelet Function Analyzer (PFA)-100 (Dade Behring AG, Düdingen, Switzerland) for drug-induced platelet dysfunction and analyzed potential contributors to HTPR with practical PFA-100 data over six years.

METHODS

We used data from 6,957 patients who underwent PFA-100 testing after receiving aspirin, clopidogrel, or non-steroidal anti-inflammatory drugs (NSAIDs). Of these, 6,163 patients were tested with only the collagen/epinephrine cartridge (Col/EPI) of PFA-100; 794 were tested with both Col/EPI and the collagen/ADP cartridge (Col/ADP). We calculated PFA-100 closure time (CT) for each drug and compared the clinical and laboratory characteristics of the patients with prolonged CTs and normal CTs (i.e., HTPR).

RESULTS

In Col/EPI, 73.2% (365/499), 72.6% (390/537), and 55.3% (3,442/6,228) patients showed prolonged CTs for aspirin, clopidogrel, and NSAIDs, respectively. In Col/ADP, prolonged CTs were observed in 37.4% (34/91), 43.2% (35/81), and 29.6% (200/676) of patients receiving aspirin, clopidogrel, and NSAIDs, respectively. Of the patients tested with both cartridges, 88.9% (48/54), 95.3% (41/43), and 89.0% (577/648) of the patients receiving aspirin, clopidogrel, and NSAIDs had prolonged CTs, and 10.0% (79/794) showed normal CTs regardless of drugs. For clopidogrel users (both cartridges), there were more patients with malignancies in the normal CT than prolonged CT group.

CONCLUSIONS

PFA-100 is not sufficiently effective for laboratory screening of drug-induced platelet dysfunction. Malignancy may contribute to clopidogrel-related HTPR in PFA-100.

Для цитирования: Kweon Oh Joo, Lim Yong Kwan, Kim Bohyun, Lee Mi Kyung, Kim Hye Ryoun. Effectiveness of Platelet Function Analyzer-100 for Laboratory Detection of Anti-Platelet Drug-Induced Platelet Dysfunction.. Annals of laboratory medicine. 2019. DOI: 10.3343/alm.2019.39.1.23.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.