DOI 10.1016/j.jcv.2014.11.017

Медицинская библиотека · Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology · 2014

Comparison of the Roche Cobas(®) 4800 HPV assay to Digene Hybrid Capture 2, Roche Linear Array and Roche Amplicor for Detection of High-Risk Human Papillomavirus Genotypes in Women undergoing treatment for cervical dysplasia.

Phillips Samuel, Garland Suzanne M, Tan Jeffery H, Quinn Michael A, Tabrizi Sepehr N

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The recently FDA (U.S. food and drug administration) approved Roche Cobas(®) 4800 (Cobas) human papillomavirus (HPV) has limited performance data compared to current HPV detection methods for test of cure in women undergoing treatment for high grade lesions.

OBJECTIVE

Evaluation of Cobas HPV assay using historical samples from women undergoing treatment for cervical dysplasia.

STUDY

DESIGN

A selection of 407 samples was tested on the Cobas assay and compared to previous results from Hybrid Capture 2, HPV Amplicor and Roche Linear Array.

RESULTS

Overall, a correlation between high-risk HPV positivity and high grade histological diagnosis was 90.6% by the Cobas, 86.1% by Hybrid Capture 2, 92.9% by HPV Amplicor and 91.8% by Roche Linear Array.

CONCLUSION

The Cobas HPV assay is comparative to both the HPV Amplicor and Roche Linear Array assays and better than Hybrid capture 2 assay in the detection of High-Risk HPV in women undergoing treatment for cervical dysplasia.

Для цитирования: Phillips Samuel, Garland Suzanne M, Tan Jeffery H, Quinn Michael A, Tabrizi Sepehr N. Comparison of the Roche Cobas(®) 4800 HPV assay to Digene Hybrid Capture 2, Roche Linear Array and Roche Amplicor for Detection of High-Risk Human Papillomavirus Genotypes in Women undergoing treatment for cervical dysplasia.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2014. DOI: 10.1016/j.jcv.2014.11.017.