MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1161/circinterventions.111.965616

Медицинская библиотека · Circulation. Cardiovascular interventions · 2011

Stent fracture, valve dysfunction, and right ventricular outflow tract reintervention after transcatheter pulmonary valve implantation: patient-related and procedural risk factors in the US Melody Valve Trial.

McElhinney Doff B, Cheatham John P, Jones Thomas K, Lock James E, Vincent Julie A, Zahn Evan M, Hellenbrand William E

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Among patients undergoing transcatheter pulmonary valve (TPV) replacement with the Melody valve, risk factors for Melody stent fracture (MSF) and right ventricular outflow tract (RVOT) reintervention have not been well defined.

METHODS AND

RESULTS

From January 2007 to January 2010, 150 patients (median age, 19 years) underwent TPV implantation in the Melody valve Investigational Device Exemption trial. Existing conduit stents from a prior catheterization were present in 37 patients (25%, fractured in 12); 1 or more new prestents were placed at the TPV implant catheterization in 51 patients. During follow-up (median, 30 months), MSF was diagnosed in 39 patients. Freedom from a diagnosis of MSF was 77±4% at 14 months (after the 1-year evaluation window) and 60±9% at 39 months (3-year window). On multivariable analysis, implant within an existing stent, new prestent, or bioprosthetic valve (combined variable) was associated with longer freedom from MSF (P<0.001), whereas TPV compression (P=0.01) and apposition to the anterior chest wall (P=0.02) were associated with shorter freedom from MSF. Freedom from RVOT reintervention was 86±4% at 27 months. Among patients with a MSF, freedom from RVOT reintervention after MSF diagnosis was 49±10% at 2 years. Factors associated with reintervention were similar to those for MSF.

CONCLUSIONS

MSF was common after TPV implant in this multicenter experience and was more likely in patients with severely obstructed RVOT conduits and when the TPV was directly behind the anterior chest wall and/or clearly compressed. A TPV implant site protected by a prestent or bioprosthetic valve was associated with lower risk of MSF and reintervention.

Для цитирования: McElhinney Doff B, Cheatham John P, Jones Thomas K, Lock James E, Vincent Julie A, Zahn Evan M, Hellenbrand William E. Stent fracture, valve dysfunction, and right ventricular outflow tract reintervention after transcatheter pulmonary valve implantation: patient-related and procedural risk factors in the US Melody Valve Trial.. Circulation. Cardiovascular interventions. 2011. DOI: 10.1161/circinterventions.111.965616.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.